Skip to main content

Platform

RAMAN

A configurable Raman imaging platform supporting multiple excitation wavelengths to address a range of spectroscopic applications. Designed for detailed chemical identification, material characterization, and spatially resolved analysis.

Configuration

Spectral range
532–596 nm
Spatial channels
1000
Wavenumber range
0–2000 cm⁻¹
RAMAN — 532 RAMAN

Platform specifications

Select a configuration above to view its technical parameters from the platform documentation.

Optical

Spectral range
532 to 596 nm
Wavenumber range
0 to 2000 cm⁻¹
Dispersion
9.5 nm/mm
Spectral resolution
10 cm⁻¹
Spatial channels
1000
Image size
8 mm × 8 mm
F#
1.2
Magnification
1
Slit size
9 mm × 0.03 mm
Optical input
Telecentric
VPH grating efficiency
85% @ 550 nm

Aberrations

Smile
12 pixels
Keystone
2 pixels

Mechanical

Size
60 mm × 60 mm × 190 mm
Lens / camera mount
C-mount

Camera

Model
Nuvu HNu 512 Gamma
Array size
512 × 512
Pixel
16 μm
Frame rate
63 fps full frame
Data output
Camera Link

Built around your application

Start with this platform. We can adapt the optics, detector, enclosure, electronics, software, and branding to your needs.

Optics

Detectors

Mechanical

Electronics

Software

OEM

Know what you need?

Talk to an engineer

Tell us your technical requirements and we'll help determine the right platform and configuration.

Still figuring it out?

Get a free hyperspectral assessment

Tell us what you're trying to measure, what your samples look like, and how you need the data delivered. We'll recommend a path forward.